TOC ANALYZER WITH SSM 5000 A (FOR SOLID SAMPLE MEASUREMENT)

  • Saturday 23 June 2012
  • Er.Bhupendra Singh
  • TOC Analyzers

    SSM-5000A
    Combining the SSM-5000A with a TOC-Vcs/cp analyzer permits analysis of many solid samples in addition to aqueous samples, including soil, sludge, and sediments. Carbon measurement in attached residues using the swab method can be conducted for cleaning validation.

    Features

    • Analysis of 1 g samples with up to 30 mg carbon content reduces weighing errors or errors due to the uneven distribution of the sample carbon content.
    • The use of unique tandem measurement cells, combining a long cell to measure aqueous samples and a short cell for solid samples, permits a single TOC analyzer to handle a range from several 10 ppb TOC in ultrapure water to 10% TOC in solid samples.
    • Measurement of inorganic carbon (carbonate) in solid samples is also easy.
    • Aqueous samples containing large amounts of suspended matter can be analyzed by inputting the sample in a sample boat.
    • Switching between TOC-Vcs/cp analysis of aqueous samples and SSM-5000A measurements of solid samples is achieved by simple screen settings.
    • The SSM-5000A can be attached to a TOC-Vcs/cp unit with Autosampler installed. (Automatic measurement of solid samples is not possible.)

    Specifications

    AnalyteTC, IC, TOC (TC-IC)
    MethodTC: Catalytically aided combustion oxidation at 900°C
    IC: Pre-acidification, oven temperature: 250°C
    Measuring rangeTC: 0.1 mg to 30 mg carbon
    IC: 0.1 mg to 20 mg carbon
    Sample Amount1 gram - aqueous content < 0.5 g
    RepeatabilityS.D. ±1% of full scale range.
    Analysis Time5 to 6 minutes at a gas rate of 500 mL/min.
    Carrier Gas99.9% O2 at 500 mL/min.
    Ambient temperature requirements5° to 35° C
    Power requirements100-127V or 220-240V as ordered. 12A: TOC-5A: SSM-7A. 50/60Hz
    Weightabout 30 kg
    *Will vary with sample type and measurement conditions
    Special Accessories Part Number Remarks
    Tandem Cell SetP/N 638-93097-00 Required to upgrade existing TOC-5000/5050 with 200mm cell.
    Connection SetP/N 638-70292-00 Only one necessary for all option boards.
    Flow UnitP/N 638-92098-02 Carrier/combustion gas controller for 500 mL/min.
    The SSM-5000 requires:
    • ROM Version 3.00 or later, P/N 638-64712-12 or -11 (TOC-5000/5050 respectively).
    • Main Board Assembly
    • Main Board Installation Plate, P/N 631-26518-00, necessary for TOC-5000/5050 with Serial Numbers prior to 30O27441.
    • Tandem Cell Set, P/N 638-93097-00, necessary to upgrade existing TOC-5000/5050.
    • Flow Unit, P/N 638-92098-02, or other means to supply carrier/combustion gas at 500 mL/min.